Sign In | Join Free | My bushorchimp.com
China Guangzhou HongCe Equipment Co., Ltd. logo
Guangzhou HongCe Equipment Co., Ltd.
We provide you with industry leading test equipment, advanced online system and solutions–anywhere in the world.
Verified Supplier

12 Years

Home > Flammability Testing Equipment >

Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent

Guangzhou HongCe Equipment Co., Ltd.
Trust Seal
Verified Supplier
Credit Check
Supplier Assessment
Contact Now

Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent

Brand Name : HongCe

Model Number : NF-1

Certification : calibration certificate (cost additional)

Place of Origin : China

MOQ : 1 set

Payment Terms : T/T

Supply Ability : 10 set /month

Delivery Time : 30 days

Packaging Details : plywood

Price : Negotiatable

Standard : IEC60695-11-5

Flame application angle : 0°, 20°, 45°

Height of flame : 12±1mm

copper block : 0.58±0.01g

Thermocouple : NiCrφ0.5 armored ±0.1%

Temperature sensing delay : 1.7S

Warranty : 1 Year

Product name : Needle - Flame Tester

Contact Now

Electrical Control Needle-Flame Tester Of Button Operation With Air Vent And Temperature Of 01100

Product Introduction

The test machine is designed and manufactured according to the requirements of the relevant standards IEC60695-11-5, the needle-flame tester is the special instrument used for the fire hazard test of the small flame caused by the fault conditions in the simulating technology assessment equipment. It is applicable to the measurement of the fire hazard for the electrical and electronic equipments and their components and parts, and also applicable to the measurement of the fire hazard for the solid electrical insulating materials or other flammable materials.

This instrument has the advanced design, exquisite material, careful workmanship, outstanding quality, stable performance, durability and artistic shape. The flame calibration and timing are automated so as to achieve the simple operation, higher accuracy and increased test efficiency.

Technical Parameters

Model No. NF-1
Standard IEC60695-11-5
Timing device 9999X0.1s
Power 220V, 50Hz
Flame application angle 0°, 20°, 45°
Height of flame 12±1mm
Combustion tube Height larger than 35mm, outer diameter Ф0.9mm, inner diameter Ф0.5mm

Temperature measuring

copper block

Mass before drilling: 0.58±0.01g
Temperature measurement Temperature rising time from 100±5℃ to 700±3℃ is 23S±1S
Thermocouple NiCrφ0.5 armored ±0.1%
Temperature sensing delay 1.7S
Range of temperature 0~1100℃
Thermostat 16A/D ±0.05%
Type of gas source Butane with the purity higher than 95%
Note: The gas butane is not included

Test Operation

  1. Set the “FLAME APPLICATION TIME TA”.
  2. Turn the switch of gas source to the position “ON”, and press the button “GAS CONNECTION”, then ignite the flame.
  3. Adjust the height of the flame to be 12±1mm by adjusting the “GAS FLOW RATE REGULATION”.
  4. Press the button “RIGHT RUN”, and the combustion tube moves forward, when the distance between the sample and the combustion tube reaches to the pre-set value, the flame begins and the “FLAME APPLICATION TIME TA” starts timing.
  5. After completing the flame application time Ta, the combustion tube moves back, and the “DURATION OF BURNING TB” starts timing automatically.
  6. When the flame is extinguished and the glowing combustion of the test specimen, the specified layer and/or the surrounding parts is no longer visible, press the button “PAUSE”, the “DURATION OF BURNING TB” stops timing, at this time, the display value on “DURATION OF BURNING TB” is the duration of burning Tb .
  7. Suspend: press the button “RESET”, and press the button “STOP”.
  8. Press the button “RESET”, the timers are all counted clear, and the combustion tube moves back automatically.
  9. After completing the test, turn the gas source switch to the “OFF” position in time, and then turn on the fan, and clean the combustion chamber.

Attentions

  1. To prevent the leakage of gas.
  2. To prevent the combustion and explosion accident.
  3. To prevent the inhalation of toxic gas or smoke.
  4. To prevent the damage of the toxic residual to damage the human health.
  5. To prevent the blockage of the combustion tube by the foreign object.
  6. To lubricate the moving parts regularly.
  7. In an emergency, turn off the power switch on the device panel, stop the system operation, clear the faults.
  8. Please do not change some default parameters arbitrarily!
  9. For more detail information, please refer to the user manual.
  10. If there are any doubts, please contact the supplier.

Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent

Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air VentElectrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent


Product Tags:

flammability tester

      

flame test equipment

      
Buy cheap Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent product

Electrical Control Needle - Flame Test Equipment For Flammability Testing Button Operation Air Vent Images

Inquiry Cart 0
Send your message to this supplier
 
*From:
*To: Guangzhou HongCe Equipment Co., Ltd.
*Subject:
*Message:
Characters Remaining: (0/3000)